• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B11LT
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r92r3j.Device analysis: the analysis results found that the b11lt instrument was received with the sleeve melted.This type of damage is consistent with the one produced by an energized device.The device was visually inspected and no damaged was found on the clear lens.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
The tip of the trocar was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8241789
MDR Text Key133346000
Report Number3005075853-2019-15825
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001126
UDI-Public10705036001126
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue NumberB11LT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Date Manufacturer Received01/07/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2018
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-