Model Number 102
Device Problem Adverse Event Without Identified Device or Use Problem
Event Date 12/20/2018
Event Type  Injury  
Event Description

It was reported that the patient was experienced an increase in depressive symptoms and was being treated as an inpatient. It was suspected that the patient's battery had depleted which resulted in the increase in depression; however, a battery life estimation was performed with the data available to the manufacturer which did not support an end-of-service condition. Replacement surgery is being pursued. No relevant surgical intervention is known to have occurred to date. No additional or relevant information has been received to date.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 102
Type of DeviceGENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key8246369
Report Number1644487-2019-00071
Device Sequence Number1
Product CodeMUZ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation PHYSICIAN ASSISTANT
Type of Report Initial
Report Date 03/07/2019
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received01/14/2019
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device EXPIRATION Date05/19/2007
Device MODEL Number102
Device LOT Number013315
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location Other
Date Manufacturer Received02/13/2019
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured07/05/2005
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 01/14/2019 Patient Sequence Number: 1
-
-