Catalog Number 283512
Device Problem Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device was received at service center and evaluated.The defect reported by the customer has been verified and repaired.On inspecting the device, further deficiencies were found to be impairing device function.The following repair activities were performed: motor does not turn: motor replaced.Short circuit in the motor cable - motor cable replaced.Resistance value out of specs: hand control set replaced."micro": preventive replacement of o-rings performed according to service manual the unit was cleaned and testing was completed per the service manual.The unit passed all functional tests and is fully operational.The motor not turning could be caused by the short circuit in the motor cable and the keypad values being out of range.A dhr review has been conducted and the results indicate that this batch of product was processed without any incident and therefore there is no evidence of manufacturing anomalies on the paperwork reviewed.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Event Description
It was reported by the affiliates via phone that the device has to be serviced.Patient was involved.There was no patient harm or surgical delay indicated.It was discovered post operatively that the device doesn't turn and the motor was jammed.Micro tornado hp w hand control was the device involved.The procedure has been completed using a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8266771
MDR Text Key134461309
Report Number1221934-2019-56082
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/23/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-