On 15 january 2019, it was reported from decatur memorial hospital that a duo unit was leaking from the manifold housing on cylinder 1.On 15 january 2019, burdick plumbing & heating co, inc was contacted about the cart and dispatched a service technician to be at the site.The technician arrived at the site and confirmed that there was leakage from the manifold housing on both cylinders.The technician replaced the cl500 manifold housings (part #91927 and lot code #0025590) and then verified that the unit was functioning as intended.The technician then returned the unit to service without further incident.The device was tested, inspected, and repaired.Service work order dms-49150-v4n4s2 on 15 january 2019.Although the reported event of leaking manifold housings was confirmed during the inspection of the device, based on the information available it is unknown what caused the manifold housings to leak.Therefore, based on the provided information, the specific root cause cannot be determined.The reported event was confirmed during inspection of the device and the device was noted to be functioning as intended after the manifold housings of the unit were replaced.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
|