• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER Back to Search Results
Catalog Number 228143
Device Problem Positioning Failure (1158)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/20/2017
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly.The complaint devices are not being returned, therefore are unavailable for a physical evaluation.A device history record review has been conducted to determine if there were any internal processing issues which would have contributed to the nature of the product complaint.Our results indicate that this batch of product was processed without incident and therefore there is no evidence of manufacturing anomalies on the paperwork reviewed.Further, a review into the depuy synthes mitek complaints system revealed no other complaints other than this event for this lot of devices that were released to distribution.We cannot discern a root cause for the reported failure mode.At this point in time, no corrective action is required and no further action is warranted.This file will remain receptive to any potential forthcoming information received that is pertinent and germane to this issue.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the etq complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
This is report 5 of 5 for the same event.The affiliate reported via email during a meniscus repair surgical procedure, after an easy placement of the 1st meniscus seam, the red lever could not be pushed through when the second peek patch was reloaded.Recharging was therefore not possible for mechanical reasons and the system had to be discarded.The procedure was completed with same like product with a five minute delay.It was reported that there were two omnispan guns and three needles being used together when the issue occurred.The first implant deployed fine, the recharge of the second patch was not possible as the red release handle was blocking.It was reported that the gun was exchanged with the second gun but had the same result as the first gun.There was no tissue damage due to the number of implants used.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Type of Device
ORTHOPAEDIC CERCLAGE APPLIER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key8281867
MDR Text Key136248532
Report Number1221934-2018-53564
Device Sequence Number1
Product Code GEF
UDI-Device Identifier10886705010059
UDI-Public10886705010059
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Catalogue Number228143
Device Lot Number3906272
Was Device Available for Evaluation? No
Date Manufacturer Received03/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-