MEDTRONIC, INC. REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Under-Sensing (1661); Device Sensing Problem (2917); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
This information is based entirely on journal literature.This event occurred outside the us.All information provided is included in this report.Patient information is limited due to confidentiality concerns.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.A request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: positional pseudo-alternating bundle-branch-block on implantable loop recorder electrogram indian pacing and electrophysiology journal.2018; 18(6):240-241.10.1016/j.Ipej.2018.09.001.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A journal article was reviewed that contained information regarding an implantable loop recorder (ilr).The author wrote about a case where there were alerts due to detected ventricular pauses.Additionally, the electrogram showed changes in the qrs axis compatible with alternating bundle branch block.Interrogation showed the alerts were due to intermittent under-sensing and the qrs axis changes were reproducible by positional changes of the breast.It was determined the ilr had been implanted in the breast.The device remains in use.Further follow up did not yet yield any additional information.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8283474
MDR Text Key134311183
Report Number2182208-2019-00184
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial
Report Date 01/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age87 YR
-
-