Catalog Number 284580
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/09/2017
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly. (b)(4). Performed service and repair functions. Reviewed service history. Attached box label, and electrical safety. The unit was evaluated and the reason for return (too much suction) could not be duplicated. Per (b)(4), replaced worn fingers on complete pressure adjuster (preventive maintenance) with tip replacement kit. Per (b)(4), added keyboard mask. This device was produced prior to the closure of the fms facility in nice, france and therefore will not have a dhr review preformed. Please see signed legacy fms batch review. This report is being filed from the (b)(4) as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the sales rep that during an unspecified surgical procedure, it was observed that the 4580 fms duo+ pump/shaver combo device had too much suction. There was no delay in the surgical procedure as an identical spare device was used to complete the procedure. There was patient involvement reported. There were no injuries, medical intervention or prolonged hospitalization. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Type of DeviceDISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key8286578
MDR Text Key134915265
Report Number1221934-2018-53727
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/09/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/09/2017
Was Device Evaluated by Manufacturer? No Answer Provided
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-