• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX M150; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115307
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/07/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during treatment with a prisma flex m150 set, an external fluid leak from the drain bag was observed.It was reported the drain bag connector was disconnected.The set was replaced to continue treatment and the user was in ¿good condition¿.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection of the drain bag showed the yellow valve of the drain bag with the set was disconnected.The reported condition was verified.The drain valve passed testing including: valve ability to withstand pressure, bonding strength of the valve to the drain tube (traction test) and valve ability to withstand multiple actuations over a treatment.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX M150
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu
MDR Report Key8293294
MDR Text Key134876765
Report Number8010182-2019-00025
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K080519
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2020
Device Catalogue Number115307
Device Lot Number18F2002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Date Manufacturer Received02/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberPR# 1540140
Patient Sequence Number1
-
-