• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-UNI-CELECT-PERM
Device Problems Migration or Expulsion of Device (1395); Difficult to Remove (1528); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Perforation of Vessels (2135); Blood Loss (2597); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Blank fields on this form indicate the information is unknown or unavailable.Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.(b)(4).It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
It is alleged that "[pt] received a cook celect filter on (b)(6) 2010".It is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unchanged, unknown or unavailable.Exemption number (b)(4).William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).Name and address for importer site: (b)(4).Investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating "celect: migration, vc perforation, unable to retrieve, bleeding, sob, swelling, depression-updated sfc." cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Manipulation in the area of the filter implant may cause migration or contribute to changes in the filter configuration and placement.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Unknown if the reported shortness of breath and bleeding are directly related to the filter.Unknown if the reported swelling and depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information to report at this time.
 
Event Description
Patient allegedly received an implant on (b)(6) 2010 via the common femoral vein due to deep vein thrombosis, lupus and pulmonary embolism.Patient alleges migration, vena cava perforation, device unable to be retrieved, bleeding.Patient states "filter is in a position where it cannot be removed.Doctors has indicated that it would be dangerous to remove and will not perform the surgery.Shortness of breath and leg swelling.Unable to have gastric bypass surgery due to filter issue.Unable to travel without doctors order due to blood pressure and clotting issue", depression.Attempted filter retrieval performed (b)(6) 2014.Patient has another non-cook filter implanted (2007).
 
Manufacturer Narrative
Exemption number e2016032.William cook europe aps (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi) (importer).H6) patient code:vessels, perforation of (2135)- listed in the ifu, blood loss (2597)- not listed in ifu.Device code: difficult to remove (1528)-not listed in the ifu, migration of device or device component (1395)- listed in ifu, appropriate term/code not available (3191)- device perforation- listed in ifu.G1) name and address for importer site: cook medical incorporated (cmi), 400 daniels way, bloomington, in 47404, registration no.: 3005580113.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key8296074
MDR Text Key134695735
Report Number3002808486-2019-00120
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K073374
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup,Followup
Report Date 05/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/23/2012
Device Catalogue NumberIGTCFS-65-UNI-CELECT-PERM
Device Lot NumberE2637587
Was Device Available for Evaluation? No
Distributor Facility Aware Date01/11/2019
Date Manufacturer Received04/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight129
-
-