Catalog Number SGC0301 |
Device Problems
Failure to Advance (2524); Material Split, Cut or Torn (4008)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
|
|
Event Description
|
This is filed to report a tear in the steerable guide catheter (sgc).It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4+.During advancement of the sgc, resistance was met with the patients tortuous anatomy above the common femoral artery and the sgc became torn one inch proximal to the sgc tip.The sgc was removed and replaced.A new sgc was used to continue the procedure.One clip was implanted, reducing the mr to 1.There were no adverse patient effects and no clinically significant delay in the procedure.
|
|
Manufacturer Narrative
|
Internal file number - (b)(4).The reported device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities.Additionally, a review of the complaint history identified no similar incidents reported from this lot.All available information was investigated and the reported failure to advance steerable guide catheter, resulting in the reported tear in the shaft appears to be related to patient morphology/pathology due to the reported tortuous anatomy.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|