While loading the caider into the robotic arm, a fault came up stating that the instrument was not useable.
The surgeon was the one that loaded it, removed it, and still got the error.
We then placed a different instrument into the arm and it worked fine.
The instrument did not show a red dot indicating that it was expired.
I want to assure that this is not a patient safety concern at this time.
The davinci robot does a safety check on every instrument that is loaded into one of the arms.
If anything does not test correctly, you cannot use the instrument on the patient; it won't let it enter the patient's body.
Here's the info on the instrument in question: cadiere forceps intuitive #470049, lot# n10181003 0057 ver -06.
This instrument needs to be cleaned and retested to see if it will be recognized by the davinci xi robot.
I have isolated the instrument in question and will be testing it in a non-patient related method.
All of the xi instruments on occasion have a potential to have a haze left on them after decontamination.
There is an intraop troubleshooting technique of wiping the contacts with a moist sponge that may have helped but was not utilized in this case.
I have followed up by giving the staff involved this troubleshooting tip and have requested for additional training in spd for cleaning the xi instruments.
The in service is scheduled in spd to refresh on the cleaning techniques and lubrication of the xi instruments.
|