U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK SURPASS STENT DELIVERY SYSTEM 3MMX25MM INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK SURPASS STENT DELIVERY SYSTEM 3MMX25MM INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Catalog Number M003SFD030250
Device Problem Premature Separation (4045)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/10/2019
Event Type  malfunction  
Manufacturer Narrative
Subject device is not available.
 
Event Description
It was reported that during procedure for stenting an aneurysm located in the left internal carotid artery (ica) in the petrous segment, the stent (subject device) became locked into the outer delivery catheter and partially deployed. Thus, the forward pressure stretched the distal part of the delivery catheter. There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSURPASS STENT DELIVERY SYSTEM 3MMX25MM
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key8303080
MDR Text Key134973321
Report Number3008881809-2019-00038
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received02/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/08/2019
Device Catalogue NumberM003SFD030250
Device Lot Number20934483
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-