Medivators field service engineer (fse) reported during a service visit for a facility's dsd-201 automated endoscope reprocessor (aer) that the facility was using maxicide plus 28 day solution, a high level disinfectant (hld).Maxicide plus is not intended for use with the aer, thus there is potential that endoscopes resprocessed in the aer were not adequately high-level disinfected and therefore, potential for patient harm.It is estimated that the facility was using the incorrect hld since november 2018.Medivators fse informed the facility to replace the maxicide plus with validated product for use with their aer and confirmed at the next visit that the aer was operating according to specification using the correct hld.Medivators fse also reported, during his service visit, that most of the facility hookups were in poor condition.Facility staff were informed and medivators fse noted the facility hookups were of good quality at the next visit.There have been no reports of patient harm.This complaint will continue being monitored in medivators complaint handling system.
|
Medivators field service engineer reported during a service visit for a facility's dsd-201 automated endoscope reprocessor (aer) that the facility was using maxicide plus 28 day solution, a high level disinfectant.Maxicide plus is not intended for use with the aer, thus there is potential that endoscopes reprocessed in the aer were not adequately high-level disinfected and, therefore, potential for patient harm.
|