• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190618
Device Problem Melted (1385)
Patient Problem No Patient Involvement (2645)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A user facility reported a fresenius 2008k2 machine with an unspecified component that was melted.The issue occurred prior to treatment, and it was confirmed there was no patient involvement.A fresenius mexico technician reportedly serviced the machine.Due diligence attempts were exhausted, but no additional information was provided.If additional information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.  additional information: it was confirmed that there are no parts available for return to the manufacturer.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation a fresenius mexico technician was called onsite to verify the issue; however, the technician indicated the machine was not available for investigation.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008K2 HEMODIALYSIS SYS OLC/DP, SPANISH
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8334745
MDR Text Key136118605
Report Number2937457-2019-00476
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K994267
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190618
Was Device Available for Evaluation? No
Device Age MO
Date Manufacturer Received03/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-