• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problems Gas/Air Leak (2946); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem ST Segment Elevation (2059)
Event Date 01/18/2019
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed four applications were performed with balloon catheter 2af284 lot number 39238 on the date of the event without any issues.A clinical issue (st elevation) occurred during the case.In conclusion, there is no indication of the performance of the sheath in relation to the clinical issue.The reported air ingress issue could not be confirmed through data analysis.The reported sheath 4fc12 lot number 45091 was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st elevation was observed.Nitroglycerin was administered with resolve.Additionally, a bubble was observed in the patient's angiogram, which was reported to indicate air ingress via an unknown route.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8337392
MDR Text Key136106995
Report Number3002648230-2019-00085
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/22/2020
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number45091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-