• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GENERAL ELECTRIC, CO GE 16 SLICE BRIGHTSPEED; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GENERAL ELECTRIC, CO GE 16 SLICE BRIGHTSPEED; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 717724CT
Device Problems Signal Artifact/Noise (1036); Gas/Air Leak (2946)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2019
Event Type  malfunction  
Event Description
Patient underwent unenhanced ct brain scan.Upon tilting the gantry, artifact was present.An air bubble was in the hermit tank causing the artifact.Hermit tank was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GE 16 SLICE BRIGHTSPEED
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
GENERAL ELECTRIC, CO
MDR Report Key8339534
MDR Text Key136885153
Report NumberMW5084067
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number717724CT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age58 YR
-
-