ARROW INTERNATIONAL INC. ARROW PI CVC KIT: 3-L 7 FR X 20 CM AGB; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CDC-45703-XP1A
Device Problems Difficult to Advance (2920); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reports the wires have been kinking and difficult to thread.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The customer returned a guide wire and straightener tube for evaluation.The guide wire was observed to have a three distinct kinks throughout the guide wire body.The distal j-bend was intact.Microscopic examination confirmed the kinks in the guide wire body.Both welds were present and were observed to be full and spherical.The kink in the guide wire were located 28mm, 168mm and 243mm from the distal tip.The overall length and outer diameter of the guide wire were measured and were found to be within specification.The guide wire was advanced through an inventory ars and needle to functionally check the guide wire.The guide wire passed through all components with minimum resistance.A manual tug test confirmed that both the distal and proximal welds were intact.A device history record review was performed on the guide wire and no relevant manufacturing issues were identified.The instructions-for-use (ifu) provided with the kit describes suggested techniques to minimize the likelihood of guide wire damage during use.The instructions caution that withdrawing the guide wire against the needle bevel or use of excessive force during removal could damage or break the wire.The report that the guide wire kinked during use was confirmed through examination of the returned sample.The guide wire was kinked throughout the guide wire body.The returned guide wire met all relevant dimensional requirements and a device history record review did not identify any manufacturing related issues.Based on the condition of the guide wire and the report that the damage was observed during use, it was determined that unintentional use error caused or contributed to this event.Teleflex will continue to monitor and trend for reports of this nature.
 
Event Description
The customer reports the wires have been kinking and difficult to thread.The patient's condition is reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PI CVC KIT: 3-L 7 FR X 20 CM AGB
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8339738
MDR Text Key136158302
Report Number1036844-2019-00125
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
K071538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date06/30/2019
Device Catalogue NumberCDC-45703-XP1A
Device Lot Number23F18E0147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2019
Date Manufacturer Received03/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-