AV-TEMECULA-CT MITRACLIP SYSTEM STEERABLE GUIDE CATHETER Back to Search Results
Catalog Number SGC0301
Device Problems Positioning Failure (1158); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/24/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The customer reported the device was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
This is filed to report the inability to straighten the steerable guide catheter (sgc) tip.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4.The sgc was inserted into the anatomy.When rotating the +/- knob in the minus direction, a pop sound was heard, and the sgc tip would not straighten.The sgc was removed and replaced with a new sgc.Two clips were implanted, reducing mr to 2.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.Additionally, a review of the complaint history identified no other incidents reported from this lot.A definitive cause for a reported unable to straighten the steerable guide catheter (sgc) could not be determined in this incident.Additionally, the reported noise appears to be a cascading effect of mechanical issue.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
STEERABLE GUIDE CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8339802
MDR Text Key136162175
Report Number2024168-2019-01108
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/08/2019
Device Catalogue NumberSGC0301
Device Lot Number81207U242
Was Device Available for Evaluation? No
Date Manufacturer Received04/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-