• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL OPTEASE RETR FILTER 55 FEMORAL; FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL OPTEASE RETR FILTER 55 FEMORAL; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number 466F220AF
Device Problem Material Frayed (1262)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2019
Event Type  malfunction  
Manufacturer Narrative
A review of the manufacturing documentation associated with lot 17805598 presented no issues during the manufacturing process that can be related to the reported event.The device was not returned for analysis.A device history record (dhr) review was performed and showed that these lots of products met all the requirements per the applicable manufacturing quality plan.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported, a sheath and the dilator of an optease were attempted to be inserted, however the distal end of the sheath became frayed.There was no reported patient injury.Therefore it was replaced with a new optease and it could be inserted.An approach was made from the femoral artery.The device will not be returned for analysis because of an emergent procedure.
 
Manufacturer Narrative
As reported, a sheath and the dilator of an optease inferior vena cava filter were attempted to be inserted, however the distal end of the sheath became frayed.There was no reported patient injury.Therefore it was replaced with a new optease and it could be inserted.An approach was made from the femoral artery.The device will not be returned for analysis because of an emergent procedure.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.Without the return of the product for analysis, the event reported by the customer as ¿brite tip (csi/filters)- frayed/split/torn¿ could not be confirmed.Procedural factors, (such as interaction of the cannula sheath with a concomitant vessel dilator or guide wire) may contribute to the failure as reported.According to the instructions for use (ifu), ¿a vessel dilator facilitates the percutaneous entry of the csi by forming an atraumatic transition from the skin through the subcutaneous tissue to the vessel.If increased resistance is felt upon insertion of the csi, investigate the cause before continuing.If the cause of the resistance cannot be determined and corrected, discontinue the procedure and withdraw the csi.¿ based on the device history record review, there is no indication that the event is related to the device manufacturing process.Therefore, no corrective or preventative actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTEASE RETR FILTER 55 FEMORAL
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI 
MDR Report Key8340302
MDR Text Key136170060
Report Number9616099-2019-02709
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Model Number466F220AF
Device Catalogue Number466F220AF
Device Lot Number17805598
Was Device Available for Evaluation? No
Date Manufacturer Received02/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-