Device evaluated by mfr: returned product consisted of a nc emerge balloon catheter with a 0.014" guidewire stuck in the guidewire lumen.Visual examination revealed a kink on the hypotube 85.5cm from the hub, the hypotube is twisted, buckling in the guidewire lumen 30mm and 37mm from the tip.There is contrast present in the inflation lumen and balloon and the balloon is loosely folded.The guidewire was attempted to be removed but was unsuccessful.Inspection of the remainder of the device presented no other damage or irregularities.
|