Catalog Number 209999
Device Problems Computer Software Problem (1112); Output Problem (3005)
Patient Problems Tissue Damage (2104); No Known Impact Or Consequence To Patient (2692)
Event Date 02/08/2019
Event Type  Injury  
Manufacturer Narrative
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical. A supplemental report will be submitted when additional information becomes available.
 
Event Description
After passing femur and cutter checkpoint, surgeon noticed burr was burring much deeper than planned. Can provide full detail on event asap. Case type: pka. Surgical delay: 30 minutes. Update per mps: the plan was to conduct a primary pka but burring was much deeper than planned and a tka was implanted instead of the pka. The surgeon did not have any issues with the burring device or the anspach motor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name3.0 RIO® ROBOTIC ARM - MICS
Type of DeviceSTEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
bethany hinson
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key8354694
MDR Text Key136653968
Report Number3005985723-2019-00161
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number209999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/05/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 02/20/2019 Patient Sequence Number: 1
-
-