• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIG FEM ADPT TORQUE WRENCH; KNEE INSTRUMENT : TORQUE DEVICES Back to Search Results
Catalog Number 961673
Device Problem Component or Accessory Incompatibility (2897)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/08/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported non-functional problem.The black plastic protector component was loose.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.The complaint sample consisted of (1) 961673 sig fem adpt torque wrench.Examination of the returned device found the black plastic protector component was loose and still attached.The loose black plastic protector could contribute to the reported "inability to tighten the implant".A search of the complaint database found additional reports of cracking, loose and missing black plastic protector.(b)(4) has been initiated to remove the radel® socket cap altogether and replace with a teflon® (ptfe ¿ polytetrafluoroethylene) split ring and a peek (polyetheretherketone) disc to the torque wrench.Based on a similar evaluation, the root cause is attributed to the device design.The current complaint sample was manufactured prior to release of products manufactured to (b)94).No further corrective action required.Complaint trends will be monitored by post market surveillance through (b)(4).Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The black plastic protector component was loose and still attached.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the returned device confirmed the reported non-functional problem.The black plastic protector component was loose.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIG FEM ADPT TORQUE WRENCH
Type of Device
KNEE INSTRUMENT : TORQUE DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8355183
MDR Text Key136779552
Report Number1818910-2019-85258
Device Sequence Number1
Product Code HXC
UDI-Device Identifier10603295239789
UDI-Public10603295239789
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 02/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number961673
Device Lot NumberA1107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Date Manufacturer Received02/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-