The generator was received and product analysis was completed.Proper device function was verified in the pulse generator¿s ability to provide appropriate programmed output currents.In the product analysis lab, the device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.No additional or relevant information has been received to date.
|