Model Number 103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Seizures (2063)
Event Date 01/26/2019
Event Type  Injury  
Event Description
It was reported that the patient was hospitalized due to an increase in seizures and status epilepticus.The reason for the increase in seizures is unknown.Clinic notes were received and reviewed for the patient¿s generator replacement.Programming data was reviewed.The battery life calculation does not support the increase in seizures being due to low battery.No anomalies were seen.The patient¿s generator was replaced, and has not been received by livanova.No other information has been received.
 
Event Description
The generator was received and product analysis was completed.Proper device function was verified in the pulse generator¿s ability to provide appropriate programmed output currents.In the product analysis lab, the device output signal was monitored for more than 24-hrs, while the pulse generator was placed in a simulated body temperature environment.Results showed no signs of variation in the pulse generator¿s output signal and demonstrated that the device provided the expected level of output current for the entire monitoring period.In addition, a comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.No additional or relevant information has been received to date.
 
Event Description
Information was received from the physician that they had hoped a new generator would help this patient as they had their device for a while.The device has not been received by livanova to date.No additional or relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8359286
MDR Text Key136790015
Report Number1644487-2019-00327
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/09/2013
Device Model Number103
Device Lot Number3131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2019
Is the Reporter a Health Professional? Yes
Device Age6 YR
Event Location Other
Date Manufacturer Received04/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age20 YR
-
-