ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B12SRT
Device Problems Material Puncture/Hole (1504); Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/30/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing record evaluation could not be performed.
 
Event Description
It was reported that during a laparoscopic urological procedure, the duckbill valve got a hole during use, and the piece of valve fell off.The fallen piece was retrieved, and no pieces were left inside the patient.There were no missing pieces of broken valve.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # r94y33 device analysis: the analysis results found that the b12srt device was returned with the duckbill damaged and torn; the torn piece of the duckbill was returned inside a petri dish.In addition, the tyvek was returned along with the instrument.The device was disassembled in order to evaluate the condition of the duckbill and the damage was confirmed; in addition, the separate piece of the duckbill was observed to have a mark which suggest that a pointy instrument was inserted through the trocar with excessive force.Per instructions for use: "use caution when introducing or removing instruments through the trocar sleeve in order to prevent inadvertent damage to the seals which could result in loss of pneumoperitoneum.Special care should be used when inserting sharp or angled edged endoscopic instruments to prevent tearing the seal." a manufacturing record evaluation was performed for the finished device batch number r94y33, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8361839
MDR Text Key137259759
Report Number3005075853-2019-16781
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10705036001188
UDI-Public10705036001188
Combination Product (y/n)N
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue NumberB12SRT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2019
Date Manufacturer Received02/25/2019
Patient Sequence Number1
-
-