ETHICON ENDO-SURGERY, LLC. ENDOPATH XCEL BLADELESS TROCAR; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
|
Back to Search Results |
|
Catalog Number B12SRT |
Device Problems
Material Puncture/Hole (1504); Device Fell (4014)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing record evaluation could not be performed.
|
|
Event Description
|
It was reported that during a laparoscopic urological procedure, the duckbill valve got a hole during use, and the piece of valve fell off.The fallen piece was retrieved, and no pieces were left inside the patient.There were no missing pieces of broken valve.Another device was used to complete the case.There were no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
(b)(4).Batch # r94y33 device analysis: the analysis results found that the b12srt device was returned with the duckbill damaged and torn; the torn piece of the duckbill was returned inside a petri dish.In addition, the tyvek was returned along with the instrument.The device was disassembled in order to evaluate the condition of the duckbill and the damage was confirmed; in addition, the separate piece of the duckbill was observed to have a mark which suggest that a pointy instrument was inserted through the trocar with excessive force.Per instructions for use: "use caution when introducing or removing instruments through the trocar sleeve in order to prevent inadvertent damage to the seals which could result in loss of pneumoperitoneum.Special care should be used when inserting sharp or angled edged endoscopic instruments to prevent tearing the seal." a manufacturing record evaluation was performed for the finished device batch number r94y33, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|