DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2019 that on (b)(6) 2019 the receiver shutdown unexpectedly.No additional event or patient information is available.No product was provided for evaluation.The complaint confirmation of the receiver ceasing to function could not be determined.A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.The device was externally visually inspected and passed.The receiver failed to charge and reboot due to the receiver not turning on.The battery was replaced with a known good battery and the receiver turned on.The log was downloaded and reviewed finding an error related to the complaint.The receiver case was opened, the internal inspection passed.The allegation was confirmed.The root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
ying chen
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8367694
MDR Text Key137140356
Report Number3004753838-2019-21029
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT20649-1
Device Catalogue NumberSTK-GL-013
Device Lot Number5234982
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received06/24/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-