Catalog Number 367344 |
Device Problems
Product Quality Problem (1506); Difficult or Delayed Activation (2577)
|
Patient Problems
Exposure to Body Fluids (1745); Injury (2348)
|
Event Date 02/07/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Date of event: unknown.The date received by manufacturer has been used for this field.Batch number (8198916) was not found for material number (367344).Medical device expiration date: unknown.(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
|
|
Event Description
|
It was reported that during use of the bd vacutainer® push button blood collection set the trigger didn't work properly and failed to retract the needle.A staff member scratch herself after using butterfly needle for blood collection.
|
|
Manufacturer Narrative
|
Investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: as no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Root cause description: as there was no sample or photo available for evaluation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that during use of the bd vacutainer® push button blood collection set the trigger didn't work properly and failed to retract the needle.A staff member scratch herself after using butterfly needle for blood collection.
|
|