• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Capsular Bag Tear (2639)
Event Date 09/14/2017
Event Type  Injury  
Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.Viscoelastic was not provided.It is unknown if a qualified product was used.The product investigation could not identify a root cause.It is unknown if a qualified viscoelastic was used.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.There has been one other complaint reported in the lot number.Additional information has been requested.The manufacturer internal reference number is: (b)(4).
 
Event Description
A facility representative reported that a patient experienced a capsular tear.The lens was extracted.Additional information is not expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Type of Device
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8375670
MDR Text Key137315901
Report Number1119421-2019-00243
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberAU00T0
Device Lot Number12464473
Was Device Available for Evaluation? No
Date Manufacturer Received09/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-