Model Number 10-80-00 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not provided.Serial number is unknown.This information will be provided in a supplemental report if made available.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.Livanova deutschland manufactures the s5 roller pump.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova deutschland received a report that a s5 roller pump could not be controlled properly during procedure.There was no report of patient injury.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
The serial number was provided and is: (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a sticky connection.Liquid medicine had adhered to the knob and control panel of the machine.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|