FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT267
Device Problem Material Puncture/Hole (1504)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The rt267 infant dual heated evaqua2 breathing circuit is not sold in the usa but it is similar to a product sold in the usa.The 510(k) for that product is k103767.Method: the complaint rt267 infant dual heated evaqua2 breathing circuits was returned to fisher & paykel healthcare regional office in india and was visually inspected.Results: visual inspection revealed multiple holes in the inspiratory limbs of the rt267 circuit.The holes are all located on the peaks of the tube with a rough edges.Conclusion: it is most likely the abrasion is due to transport or storage condition.All rt267 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject rt267 circuit would have met the required specification at the time of production.Our user instructions that accompany the rt267 infant dual heated evaqua2 breathing circuit state the following: - check all connections are tight before use.- perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- set appropriate ventilator alarms.
 
Event Description
A distributor in (b)(4) reported via a fisher & paykel healthcare field representative that holes were found in the tubing of rt267 infant dual heated evaqua2 breathing circuits.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite100
irvine, CA 92618
9494534000
MDR Report Key8380582
MDR Text Key137847475
Report Number9611451-2019-00179
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT267
Device Catalogue NumberRT267
Device Lot Number2100530392
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-