Model Number BEQ-HLS 7050 USA#HL
Device Problem Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).Device requested but not received to date.
 
Event Description
According to the customer: excessive venous pressure from 100 to 180mmhg with no visible clotting, able to flow adequately however concerned pressure sensor inaccurate while supporting a (b)(6) pediatric pt.Circuit changed to new beq hls kit 7.0 circuit.No harm to the patient was reported.(b)(4).
 
Manufacturer Narrative
Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary ag kehler strasse 31, 76437 rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.Importer- (b)(4).Maquet cardiopulmonary gmbh requested the product for investigation in the laboratory.The product was returned but was not investigated in the complaints lab because the device may contain biological residue of who risk group 3 and 4 considered to be infectious such as hiv, hepatitis a or b.(see attachment investigation summary report).Therefore no laboratory investigation could be performed by the manufacturer.A review for similar complaints to be investigated already was performed and no similar complaints were found.Thus the reported failure could not be confirmed.Thus the reported failure could not be confirmed.To perform a dhr review is not possible at this time as no serial number is available.Based on the information available at this time the cause of this failure was determined to not be attributed to a device related malfunction.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
Internal reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
MDR Report Key8390063
MDR Text Key141478761
Report Number8010762-2019-00058
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Type of Report Initial,Followup
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/09/2020
Device Model NumberBEQ-HLS 7050 USA#HL
Device Catalogue Number701052794
Device Lot Number70127517
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-