It was reported that after continuous veno-venous hemofiltration treatment using a prismaflex control unit, a prismaflex set and a prismasol bag, an intensive care unit patient experienced tachycardia and a facial and toe twitch.
It was reported while the patient was connected, the blood backed up from the deaeration chamber.
After the blood was returned, a new set was primed.
Shortly after treatment started, the control unit generated a flow problem alarm.
The control unit alarmed 14 times in total.
The treatment was stopped with partial blood restitution (140 of 189 ml returned).
The total amount blood loss was estimated as 330ml.
Approximately 20 min after treatment ended, the patient experienced tachycardia, reported as ¿150s¿ and a facial and toe twitch.
No further detail was provided regarding any medical interventions for the events.
At the time of this report, the patient remained in the icu and the patient¿s outcome was not reported.
No additional information is available.
|