Model Number 24970A |
Device Problems
Failure to Capture (1081); High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the mobile programmer failed to capture, during lead analysis.It was further reported that impedance measurements were high and undefined.It was noted that the patient connector cable was changed, and leads were repositioned, but the issue persisted.Lead analysis was completed successfully, with a different model analyzer.The programmer remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Failure analysis: could not confirm customer complaint.No defect found.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product analysis: the programmer base's connector port was found to have broken pins and damaged barrels.The base was forwarded for further analysis.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
(b)(6) 2019: the programmer base was returned for analysis.
|
|