MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL TCK1 HD CAMERA HEAD; ENDOSCOPIC VIDEO CAMERA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL TCK1 HD CAMERA HEAD; ENDOSCOPIC VIDEO CAMERA Back to Search Results
Catalog Number 242400
Device Problem Intermittent Loss of Power (4016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
It was reported by the affiliate that thetck1 hd camera head image is flickering intermittently and color bars are appearing.The event occurred pre-operatively.The intended procedure was a turp.There was no surgical delay or patient consequences.The existing camera was used to complete the procedure; an alternate product was readily available.No surgical intervention is planned.
 
Manufacturer Narrative
Product complaint (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary :the complaint device was received at the service center and evaluated.The complaint can be confirmed since the tck1 camera head was working as intended post firmware upgrade and burning test.We cannot discern a root cause for the reported failure mode.The unit passed all functional tests and is fully operational.A manufacturing record evaluation was performed for the finished device [product code: 242400, serial number: (b)(4).And no non-conformances were identified per qlik query.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi:(b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: corrected data: g4: alert date on initial medwatch report should have been january 21, 2019.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11: corrected data: g4: awareness date reported on follow up report 1 as january 21, 2019 but should have been march 11, 2019.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TCK1 HD CAMERA HEAD
Type of Device
ENDOSCOPIC VIDEO CAMERA
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key8407826
MDR Text Key138380868
Report Number1221934-2019-56601
Device Sequence Number1
Product Code FWF
UDI-Device Identifier10886705026258
UDI-Public10886705026258
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number242400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Date Manufacturer Received10/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-