BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX HF1000; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 107140
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/13/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
This was reported that during patient treatment with a prismaflex control unit and a prismaflex hf1000 filter set, air was observed due to a loose replacement line connection.Flow problem alarms were triggered by the machine.The customer secured the connection and adjusted the chamber level.The cannot detect return and return disconnection alarms occurred.The customer noted that the effluent line was "jumping" significantly and the blood flow path of the circuit was clearing.The access line was clear, from the patient, all the way through the filter and into the deaeration chamber.It was reported several large clots were visible in the filter.The customer disconnected the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX HF1000
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
MDR Report Key8408041
MDR Text Key138571973
Report Number8010182-2019-00066
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
PMA/PMN Number
K042938
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 03/11/2019,04/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number107140
Device Lot NumberASKU
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/11/2019
Distributor Facility Aware Date02/13/2019
Event Location Hospital
Date Report to Manufacturer03/11/2019
Date Manufacturer Received03/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PRISMAFLEX MACHINE.
-
-