MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE CARDIO PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE CARDIO PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS Back to Search Results
Model Number MERGE CARDIO V10.1.2
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/12/2019
Event Type  malfunction  
Manufacturer Narrative
At this time, support is still working with this customer to investigate and correct the issue of the mpa es measurement populating the mpa ed measurement field. Mpa es values are significantly different than mpa ed values. Therefore, the issue is highly detectable for a trained/qualified cardiology user. When more information becomes available, a supplemental will be submitted.
 
Event Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data. On (b)(6) 2019, a customer reported that the main pulmonary artery (mpa) end systole (es) measurement is populating the mpa end diastole (ed) measurement field in cardio clinical reporting. Due to an incorrect values displaying in the diagnostic report, there is a potential for incorrect treatment of a patient that could result in harm. Reference complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERGE CARDIO
Type of DevicePICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge dr.
hartland, WI 53029
2629123514
MDR Report Key8408644
MDR Text Key139841318
Report Number2183926-2019-00011
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 02/12/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMERGE CARDIO V10.1.2
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/11/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-