Model Number 106
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2015
Event Type  malfunction  
Event Description
The patient's generator was replaced due to reported prophylactic battery replacement.The explanted product was received by the manufacturer at end of service condition.Product analysis was completed on the returned generator and found evidence of premature battery depletion.Upon opening the generator can, visual analysis identified contaminates on the trimmed edge of the printed circuit board assembly (pcba).The post-burn electrical test indicated that the contamination on the trimmed edge of the pcba was suggestive of probable electrical paths that were established between the copper edges, which could contribute to premature battery depletion.The manufacturer's device history records were reviewed.The generator passed final quality and functional specifications prior to release.The generator was subjected to laser-routing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
rachel kohn
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key8409943
MDR Text Key138509073
Report Number1644487-2019-00471
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/07/2017
Device Model Number106
Device Lot Number4410
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2019
Event Location Other
Date Manufacturer Received02/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age14 YR
-
-