The device has not be returned for analysis.
Since the device was not returned, we are unable to perform further root cause analysis.
All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.
If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that during use of the second pipeline flex with shield, the adjunctive coiling catheter come out of the aneurysm and it could be repositioned into the aneurysm.
However, it again prolapsed into the parent vessel along with the first coil.
An attempt was made to recapture the pipeline, but the wire separated from the device.
The pipeline separated from the resheathing pad.
All devices were removed from the patient.
No device was reported to have been implanted in the patient.
The procedure was not completed, and the patient was scheduled for retreated on (b)(6) 2019.
The patient was undergoing embolization treatment of a ruptured blister aneurysm (unknown dimensions) located in the right internal carotid artery (ica) above the pca.
The distal and proximal landing zones were unknown.
The patient¿s vasculature was normal in tortuosity.
|