Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Fistula (1862); Blood Loss (2597)
Event Date 03/13/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, when the physician removed the catheters and sheath, severe bleeding was observed at the sheath puncture site. Compression was performed without success. A vascular surgeon was called to diagnose the issue. It was concluded that the patient had a arterio-venous fistula which needed to be repaired in surgery. The fistula was repaired without further complication. It was also noted that the patient experienced phrenic nerve palsy (pnp). The case was completed with cryo. No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the pnp fully recovered. However, the patient developed an abscess a few days after the procedure and is currently hospitalized.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed. The data files showed at least 5 applications were performed with balloon catheter 2af283 with lot number 22083, without any issue on the date of the event. Clinical issues were encountered during the case. There is no indication of relation of adverse event to the performance of the cryo device. The physical product was not returned for investigation. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8422803
MDR Text Key138891838
Report Number3002648230-2019-00191
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number4FC12
Device Catalogue Number4FC12
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/16/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/14/2019 Patient Sequence Number: 1
-
-