• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2019 that on (b)(6) 2019 the receiver ceased to function.No additional event or patient information is available.No product or data was provided for evaluation.The confirmation of the complaint is undetermined.The probable cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated.An external visual inspection was performed and passed.Receiver charged and will boot was performed and failed due to call tech support displaying on the screen.Pictures were taken.Internal visual inspection was performed and failed due to water damage.Pictures were taken.The log was not downloaded for review due to the receiver not communicating.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key8433370
MDR Text Key139249919
Report Number3004753838-2019-027256
Device Sequence Number1
Product Code MDS
UDI-Device Identifier10386270000153
UDI-Public10386270000153
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMT22495
Device Catalogue NumberSTK-DR-001
Device Lot Number5239210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient Weight100
-
-