DEPUY ORTHOPAEDICS, INC. 1818910 SELF CENT HIP 54X28 BLU; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
|
Back to Search Results |
|
Catalog Number 103554000 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
Not Applicable (3189)
|
Event Date 03/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
|
|
Event Description
|
It was reported that the plastic portion of bipolar head would not fully engage in metal portion.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: update may 08, 2019.The device has been received for examination.Examination of the returned device is not able to confirm the reported problem.The components can be assembled and no issue with fit or function is identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected: h3.Product complaint # = > (b)(4).Investigation summary = > update may 08, 2019.The device has been received for examination.Examination of the returned device is not able to confirm the reported problem.The components can be assembled and no issue with fit or function is identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot = > null.Device history batch = > null.Device history review = > null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|