Model Number 60-02-15 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/22/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova (b)(4) received a report that the encoder of a centrifugal pump system with tubing clamp would not control the flow speed.There was no patient involvement.
|
|
Manufacturer Narrative
|
Replaced encoder (b)(4) and tested unit.All functions works properly.Perfusion was satisfied with the repair and returned the unit back into service.A review of the dhr did not identify any deviations or nonconformities relevant to the issue.
|
|
Event Description
|
See initial report.
|
|