Further information was received confirming that electrocautery was not used on the patient at time of implant or since device implant.Generator product analysis was completed.The product analysis lab confirmed that the generator was at premature end of life ifi(intensified follow-up indicator)=yes.A visual assessment on the pcba showed no visual anomalies.A number of test parameters, related to ¿supply current¿, were found to be out of specification.Based on the postburn electrical test results, the ¿supply current¿ test measurements demonstrate an increased current consumption for the pulse generator.The capacitor¿s (c1) increase in leakage current resulted in increased current consumption (out of specification) for both standby and pulsing modes of operation, and may have been the contributing factor for the reported allegation of premature end of life.The cause for the capacitor¿s (c1) increase in leakage current could not be determined.There were no other performance of any other type of adverse events found with the pulse generator.
|