• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Catalog Number IAP-0700
Device Problem Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/27/2019
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
It was reported that the intra-aortic balloon pump (iabp) while in use event on a patient the pump was powered up in the locked screen mode, it was noted that it could not be unlock with key press. Calibrating the screen with the up/down arrows corrected the problem. The pump was swapped out for another. As a result, the display was replaced. There was no report of patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAC3 OPTIMUS IABP NA/EMEA
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
carmen sherman
16 elizabeth drive
chelmsford, MA 01824
9782505100
MDR Report Key8445663
MDR Text Key139682652
Report Number3010532612-2019-00073
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902084965
UDI-Public00801902084965
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/27/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/26/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-