• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550225-23
Device Problems Difficult to Insert (1316); Leak/Splash (1354); Product Quality Problem (1506); Activation Failure (3270)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/04/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that the 2.25x23 mm xience sierra stent delivery system (sds) was difficult to load on the guide wire.When it was advanced to the distal left anterior descending coronary artery, it did not feel right.Attempt was made to deploy the xience at the lesion, but it would not deploy.After the sds was removed from the anatomy, attempt was made to deploy it and contrast was noted coming out the distal end of the hypotube.Another same size xience sierra was used to complete the procedure without further incident.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Manufacturer Narrative
Internal file number - (b)(4).Evaluation summary: the device was returned and analysis was performed.The reported physical property issue, leak and failure to deploy were confirmed.The reported difficulty to insert was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key8449394
MDR Text Key139750955
Report Number2024168-2019-02298
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/19/2019
Device Catalogue Number1550225-23
Device Lot Number8091341
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/08/2019
Date Manufacturer Received03/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-