DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-1
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/28/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2019, that on (b)(6) 2019, that an unexpected receiver shutdown occured. No additional event or patient information is available. No product or data was provided for evaluation. The complaint confirmation of the reported unexpected receiver shutdown could not be determined. A root cause could not be determined.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated. The device was externally visually inspected and it passed. The receiver failed to charge and reboot. The receiver did not turn on. The battery was replaced with a known good battery and the receiver turned on. The log was downloaded and reviewed finding an error related to the complaint. The receiver case was opened, the internal inspection failed with a damaged battery. The allegation was confirmed. The root cause was determined to be a damaged battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key8449715
MDR Text Key139762361
Report Number3004753838-2019-28837
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMT20649-1
Device Catalogue NumberSTR-GL-004
Device Lot Number5224380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-