Pfs alleges injuries, pain, muscle spasm, walking difficulty,sleep disturbance, pseudotumors, metallosis and itching around scar tissue.
After review of medical records, the patient was revised to address metallosis and pseudotumor resulting to pain.
Operative finding reported large pseudotumor and pseudocapsule, a large amount of metal stain fluid present, pseudotumor was dissected, also had a large amount of metal staining around greater trochanter associated with synovitis and bursitis, scar tissue were debrided, some heterotropic bone proximal to the acetabular component, and there was a large amount of metal debris around the taper.
There was also corrosion on the femoral head and trunnion.
However, trunnion was intact and was not revised.
Lab result shows metal ion levels were below 7ppb.
Doi: (b)(6) 2008; dor: (b)(6) 2016: left hip.
|
Product complaint # (b)(4).
Investigation summary == > no device associated with this report was received for examination.
The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.
Depuy considers the investigation closed.
Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|