• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 8022 HAND PC TORNADO SHAVER -NS; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 8022 HAND PC TORNADO SHAVER -NS; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Catalog Number 288022
Device Problem Incomplete or Inadequate Connection (4037)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the device was received at the service center and evaluated.The device was received for maintenance and there were defects impairing the function of the device.The following repair activities were performed: "tornado": preventive replacement of o-rings performed.Motor does not turn: motor replaced.Motor corroded - motor replaced.Damaged parts of the drill mounting mechanism replaced.Defective o-rings replaced.Fluid ingress into the system is the possible root cause for the corroded motor and the motor not functioning as intended.We cannot determine a root cause for the damaged drill mounting mechanism.The unit was cleaned and the testing of the unit was completed per the service manual.The unit passed all functional tests and is fully operational.A manufacturing record evaluation was performed for the finished device [product code: 288022, serial number: (b)(4)] , and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).
 
Event Description
It was reported via service request that the micro tornado hand piece with hand controls needs maintenance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8022 HAND PC TORNADO SHAVER -NS
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8451513
MDR Text Key139837245
Report Number1221934-2019-56689
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705018345
UDI-Public10886705018345
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Type of Report Initial
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number288022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2019
Date Manufacturer Received03/25/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-