Model Number 50000000
Device Problems Break (1069); Display or Visual Feedback Problem (1184); Leak/Splash (1354); Defective Component (2292); Failure to Calibrate (2440)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device experienced flow rate issues. During evaluation with technical support, the drain port broke and began to leak. Per additional information received from biomed (b)(6) via technical support on (b)(6) 2019, the drain valves were replaced. He attempted to calibrate the device, and the device failed calibration with an error 80 (not heating). He confirmed that the heater failed by using a multi-meter. The device hours was noted at 1786.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key8453841
MDR Text Key140124131
Report Number1018233-2019-01550
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 05/28/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received03/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-