According to the reporter, during a laparoscopic inguinal hernia repair, while closing the peritoneum, the staples would not form so the device was pulled out but got caught on the peritoneum and the surgeon had to cut the peritoneum to take the device out.Another device of the same kind was used.
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual inspection of the instrument noted the e-piece was disengaged and missing from the device.No device loading unit (dlu) was received.Functional testing was precluded due to the disengaged e piece.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur as a result of excessive manipulation of the device.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|