No patient information provided as no patient was involved in this concern.
A medtronic representative went to the site to test the equipment.
Testing revealed that after the cable was replaced, the issue was resolved.
The system then passed the system checkout and was found to be fully functional.
The power cable was returned to the manufacturer for analysis.
Analysis found that the ground prong was broken off and was missing.
Analysis found that the reported event was related to a mechanical issue.
If information is provided in the future, a supplemental report will be issued.
|